Thermocouple Compensating Lead Wire

Thermocouple Compensating Lead Wire
Thermocouple Compensating Lead Wire
Request a Quote

Thermocouple compensating lead wires are used to connect the electrode (cold side of a thermocouple) with display instruments in order to complete a readable temperature measurement system. The working temperature of the wire is -25~200℃. Our plant produces a variety of these wires in order to meet your specific needs. They comply with both Chinese and international thermocouple standards.

Note:
1. Do not store for longer than six months.
2. Do not use in humid environments.

Properties:
EMF of Compensating Wire and the Tolerance

Thermocouple No. indexing Compensation wire type Usag Classification Measuring junction temperature / ℃ standard EMF value/μV Precision grade General grade
Tolerance μV Thermal EMF range μV Tolerance μV Thermal EMF range μV
S or R SC or RC G 100 646 ±30 616~676 ±60 586~706
H 100
200
646
1 1441
-
-
-
-
±60 586~706
1.381~1 501
K KX,KCA
or KCB
G -25
100
-968
4 096
±40 -924~-1 012
4 052~4 140
±80 -880~-1 056
48~4 184
H -25
100
200
-968
4 096
8 138
±40 -924~-1 012
4 052~4 140
8 094~8 182
±88 -880~-1 056
48~4 184
8 050~8 226
N NX or NC G -25
100
-646
2 774
±43 -603~-689
2 731~2 817
±86 -560~-732
2 688~2 860
H -25
100
200
-646
2 774
5 913
±43 -603~-690
2 731~2 817
5 870~5 956
±86 -560~-732
2 688~2 860
5 827~5 999
E EX G -25
100
-1 432
6 319
±81 -1 351~-1 513
6 238~6400
±138 -1 294~-1 570
6 181~6 457
H -25
100
200
-1 432
6 319
13 421
±81 -1 351~- 1513
6 238~6 400
13 340~13 502
±138 -1 294~-1 570
6 181~6 457
13 283~13 559
J JX G -25
100
-1 239
5 269
±62 -1 177~-1 301
5 207~5 331
±123 -1 116~-1 362
5 146~5 392
H -25
100
200
-1 239
5 269
10 779
±62 - 177~-1 301
5 207~5 331
10 717~10 841
±123 -1 116~-1 362
5 146~5 392
10 656~10 902
T TX G -25
100
-940
4 279
±30 -910~-970
4 249~4 309
±60 -850~-1
4 189~4 339
H -25
100
200
-940
4 279
9 288
±48 -892~-988
4231~4 327
9 240~
±90 -850~-1 030
4 189~4 369
9 198~9 378
Related products
Send Message
Other Products
Inquiry
Inquiry