Vaccum Drying Oven (Two Shelf)

Vaccum Drying Oven (Two Shelf)
Request a Quote
Main technical parameter index(can be measured at an ambient temperature of +5 to +35 °C)
Structure
  • Working chamber size: 850mm height, 800mm width, 1600mm depth.
  • Effective working size: 590mm height, 660mm width, 1600mm depth. 2 chambers.
  • Dimensions: Refer to the schematic diagram.
  • Insulation thickness: 90mm.
  • Supporting foot: 100mm.
  • Heater position: Around the inside of the cavity.
  • Pumping port position: 2 chambers are concentrated in one suction port and vacuumed, controlled by solenoid valve
  • Venting port position: Equipped with inflatable solenoid valve.
  • Opening method: 2 single doors.
  • Control box position: Right side.
  • Silicone Rubber high temperature resistance sealing ring.
  • Equipped with water cooling cycle system.
  • Each chamber is equipped with a stainless steel trolley, a small wheel with a universal wheel, and each trolley with 2 directional wheels.
  • Each chamber has a blast wheel on the back, magnetic fluid connection.
Main Property
  • Working temperature: Room temperature +15 ° C ~ 180 ° C
  • Instrument accuracy: ±1℃
  • Temperature uniformity: ±2°C (empty box, 120°C, 2 hours after stable)
  • Heating time: ≤40min(empty box,25℃~150℃)
  • Temperature control method: PID adjustment
  • Vacuum degree: ≥100Pa (0.0999MPa)
  • average air leakage in 24 hours: ≤0.5kPa/h;
  • Outside surface temperature: ≤ room temperature +20 ° C
Electrical capacity:
  • Power supply: AC380V 3-phase 4-wire 50Hz
  • Heating power: 12KW*2
  • Blast power: 370W*2
  • Maximum current: 20A*2
  • Air switch capacity:50A
Security devices:
  • Air switch (prevent overcurrent and short circuit)
  • Overcurrent protection (prevent high current)
  • Over temperature protection (prevent high temperature, protect the product), Over temperature alarm.
Functional requirements
  • Program control, color LCD touch screen display settings
  • Equipped with 2 electromagnetic valves, and vacuuming, holding pressure, inflation processes are all program controlled and automatic.
  • Users can freely adjust the vacuum degree and temperature, and automatically maintain them under a specific temperature and vacuum degree range.
  • Equipment operation processes (vacuum, holding pressure, inflation) are automatically PLC controlled.